Precision Premium Oven 6-3
Thermo Scientific
Section 6
Service
To replace a defective door handle, perform the steps below:
1. Remove the two mounting screws holding latch cover in place.
2. Remove the two mounting screws holding defective handle in place.
3. Mount the replacement handle using two screws.
4. Adjust bottom nut (13/16) from end of shaft.
5. Secure latch cover in place with two screws.
Due to handling in shipment or to heat distortion with use, the door cam
may require adjustment. To facilitate proper closing and sealing of door,
Steps 1 through 6 may have to be performed more than once.
To adjust the door cam, perform the following:
1. Open door and remove screws holding latch cover in place.
2. Locate nuts securing tongue on cam shaft.
3. Loosen but do not remove outside nut.
4. Adjust inside nut, one full turn clockwise draws door 1/16” closer to
cabinet when door is closed.
5. Secure cam tongue in place by tightening outside nut.
6. Secure latch cover in place with two screws.
Replacing the Door
Handle
Adjusting the Door
Cam