The following table lists the available commands for monitoring or
controlling your unit.
To View Data: If Status Update Time is “0.0” no data will automatically
be displayed. To view data within a particular command code, type that
command code followed by the question mark (?) symbol and then
ENTER (Return).
Example: To view Setpoint Temperature type: T? <CR>
To Change Data: To change the value within the command type the
command code followed by the new value and ENTER (Return).
Example 1: To change Setpoint Temperature to 25.0°C type: T25.0
<CR>
Example 2: To change Setpoint Temperature to -.8°C type: T-.8<CR>
(A) Actual Temperature:
The actual temperature within the growth
chamber.
6-6
Plant Growth Chamber
Thermo Scientific
Section 6
Factory Options
Using the Communication
(continued)
View
Command:
Description:
Unit:
Range:
Change
Command
A?
Actual Temperature
Degrees
-10 to 50
Read Only
T?
Setpoint Temperature
Degrees
-10 to 50
T10 thru T50
R?
Baud Rate
-
0 - 5
R0 thru R5
S?
Status Update Time
-
0 - 8
S0 thru S8
V?
EPROM Version
Number
-
Current Rev.
Read Only
D?
Status Display Type
-
0 - 2
D0 thru D2
O?
Display the Alarm
Status
-
0 - 2
Read Onl