Orion Star A215 Benchtop pH/Conductivity Meters
11
Conductivity Temperature Compensation Selection
The temperature compensation can be turned off or set to Linear, nLFn (non-linear ultra pure non-degassed
water), nLFu (non-linear ultra pure degassed water) or EP (temperature compensation off and warning is
displayed if conductivity values are outside EP requirements for ultra pure water) and is used to report all
conductivity measurements at the selected reference temperature.
1. In the measurement mode, press
.
2. Press
,
,
or
to highlight
COND Channel
and press
f3 (select)
.
3. Press
or
to highlight
Mode and Settings
and press
f3 (select)
.
4. Press
or
to highlight
Temp. Comp
. and press
f3 (select)
.
5. Press
or
to highlight
Off, Linear, nLFn, nLFu
or
EP
and press
f3 (select)
.
6. Press
to return to the measurement mode.
Viewing the Calibration Log
1. In the measurement mode, press
.
2. Press
or
to highlight
Calibration Log
and press
f2 (accept)
.
3. Press
or
to highlight
pH – Channel
or
Conductivity – Channel
and press
f2 (select)
.
4. Press
or
to highlight
pH
,
RmV
,
ORP
or
ISE
(pH - Channel) or
Conductivity, Resistivity, TDS,
or
Salinity
(Conductivity - Channel) and press
f2 (select)
.
5. The meter will display a list of calibrations for the selected channel and calibration type. The list shows the
sequential number of the calibration and the date and time it was saved (
07/01/2011 12:45
).
6. To view the calibration data, press
or
to highlight a specific calibration and press
f2 (select)
.
Press
f2 (print)
to print the calibration, press
f3 (info)
to view the electrode slope between pH buffer
points (pH - Channel only) or press
f1 (back)
to return to the list of calibrations.
7. Press
to return to the measurement mode.