background image

22

Thermo Scientific Orion Star A112 Benchtop and Star A122 Portable Conductivity Meters

Table of Conductivity Standard Values vs. Temperature

Cat. No.

011005

011006

011007

01100910

011008

Temperature  

(°C)

111.9 mS/cm 

Conductivity 

Standard 

(mS/cm)

12.9 mS/cm 

Conductivity 

Standard 

(mS/cm)

1413 μS/cm 

Conductivity 

Standard 

(μS/cm)

147 μS/cm 

Conductivity 

Standard 

(μS/cm)

100 μS/cm 

Conductivity 

Standard 

(μS/cm)

0 65.10

7.135

776

81  54

1

66.84

7.344

799

83 56

2 68.59

7.555

822

86  58

3 70.35

7.768

846

88  59

4 72.12

7.983

870

91  61

5

73.91

8.200

894

93 63

6 75.70

8.418

918

96  64

7

77.50

8.638

943

98 66

8

79.32

8.860

968

101 68

9 81.15

9.084

992

103  70

10 82.98

9.309

1017

106  72

11 84.83

9.535

1043

108  73

12 86.69

9.763

1068

111  75

13 88.56

9.993

1094

114  77

14 90.45

10.22

1119

116  79

15 92.34

10.46

1145

119  81

16 94.24

10.69

1171

122  83

17 96.15

10.93

1198

125  85

18 98.08

11.16

1224

127  87

19 100.0

11.40

1251

130  88

20 102.0

11.64

1277

133  90

21 103.9

11.88

1304

136  92

22

105.9

12.12

1331

138 94

23

107.9

12.36

1358

141 96

24 109.9

12.61

1386

144  98

25 111.9

12.85

1413

147  100

26 113.9

13.10

1441

150  102

27

115.9

13.35

1468

153 104

28 117.9

13.59

1496

156  106

29 120.0

13.84

1524

159  108

30 122.0

14.09

1552

161  110

31 124.1

14.34

1580

164  112

32 126.2

14.59

1608

167  114

33 128.3

14.85

1636

170  117

34 130.4

15.10

1665

173  119

Содержание Orion Star A112

Страница 1: ...Thermo Scienti c Orion Star A112 Benchtop and Star A122 Portable Conductivity Meters Reference Guide...

Страница 2: ...Triode Minimum Stir Requirement MSR NISS Optimum Results Orion Dual Star Orion Star SAOB SMART AVERAGING SMART STABILITY Star LogR Star Navigator 21 Stat Face Triode are trademarks of Thermo Fisher Sc...

Страница 3: ...ration 11 Conductivity TDS Measurement 11 Temperature Measurement and Calibration 12 Temperature Measurement 12 Temperature Calibration 12 Chapter 5 Data Storage and Review 13 Automatic Datalog with A...

Страница 4: ...This page intentionally left blank...

Страница 5: ...vity meter These meters are capable of measuring conductivity in S cm or mS cm TDS in mg L and temperature in C or F The Orion Star A112 benchtop conductivity meters are IP54 rated The Orion Star A122...

Страница 6: ...This page intentionally left blank...

Страница 7: ...our AA batteries Con rm that the meter is off and remove the battery compartment cover On the Star A122 portable meter to remove the battery compartment cover i Loosen the screws ii Release the top po...

Страница 8: ...is running on AC power Orion Star A112 Conductivity benchtop meter includes adapter Shown when the meter is on AUTO READ mode Default setting AR and unit of measure will blink until the reading is sta...

Страница 9: ...ure for the type of spill In the measurement screen Press to take a measurement In the setup screen Press to escape the setup menu In the calibration screen Press to abort calibration Press to turn th...

Страница 10: ...6 Thermo Scienti c Orion Star A112 Benchtop and Star A122 Portable Conductivity Meters This page intentionally left blank...

Страница 11: ...ne 2 Press to enter into the submenu options 3 For main setup options with more than one submenu 1 0 Con guration 5 0 Datalog Clear a Press or until the desired submenu is shown b Press to enter into...

Страница 12: ...to 105 0 Manual Temperature Compensation Value This value is used when there is no ATC probe connected and can be set with the meter s temperature ranges The temperature unit for this value will matc...

Страница 13: ...ll ash When the reading has been stable READY will appear This read type is useful when performing an experiment that requires continuous measurements to be taken regardless of the measurement stabili...

Страница 14: ...10 Thermo Scienti c Orion Star A112 Benchtop and Star A122 Portable Conductivity Meters This page intentionally left blank...

Страница 15: ...AL this indicates a manual calibration mode a Press or to adjust the value to match your standard b Press to view the calculated cell constant and after 2 seconds the meter will proceed to measurement...

Страница 16: ...easurements place the probe into the next sample and repeat steps 2 and 3 4 When nished measuring all samples store probe according to probe instructions Temperature Calibration The meter temperature...

Страница 17: ...eturn to measurement mode Each time the reading is locked onto the screen with the AR icon The reading will automatically be stored in the datalog Manual Datalog with Continuous Read Mode 1 In measure...

Страница 18: ...14 Thermo Scienti c Orion Star A112 Benchtop and Star A122 Portable Conductivity Meters This page intentionally left blank...

Страница 19: ...w standards Memory is full The meter will automatically change to the submenu to clear the datalog CLR on the top line NO on the second If the existing memory s data is still needed 1 Press to return...

Страница 20: ...nductivity probe that is connected to the meter Assistance After troubleshooting all components of your measurement system contact Technical Support Within the United States call 1 800 225 1480 and ou...

Страница 21: ...rion Star A112 Conductivity Orion Star A122 Conductivity Orion Star A113 DO Orion Star A123 DO Equipment Class Measurement control and laboratory Orion Star A series meters are EMC Class A Conforms wi...

Страница 22: ...arranty 3 years meter replacement Universal Power Adapter Operating Conditions Operating Ambient Temperature 0 to 50 C Operating Relative Humidity 0 to 90 non condensing Storage Temperature 20 to 75 C...

Страница 23: ...STAR A122 Portable Conductivity Meter STARA1225 Orion STAR A122 Portable Conductivity Meter Kit STARA1230 Orion STAR A123 Portable Dissolved Oxygen Meter STARA1235 Orion STAR A123 Portable Dissolved...

Страница 24: ...20 Thermo Scienti c Orion Star A112 Benchtop and Star A122 Portable Conductivity Meters This page intentionally left blank...

Страница 25: ...0 96 98 Sulfuric Acid 2 84 Sugar Syrup 5 64 Total Dissolved Solids TDS The Orion Star A112 benchtop conductivity and Star A122 portable conductivity meters measure TDS as the total amount of dissolved...

Страница 26: ...77 50 8 638 943 98 66 8 79 32 8 860 968 101 68 9 81 15 9 084 992 103 70 10 82 98 9 309 1017 106 72 11 84 83 9 535 1043 108 73 12 86 69 9 763 1068 111 75 13 88 56 9 993 1094 114 77 14 90 45 10 22 1119...

Страница 27: ...d S cm 100 S cm Conductivity Standard S cm 35 132 5 15 35 1693 176 121 36 134 6 15 61 1722 179 123 37 136 7 15 86 1751 182 125 38 138 9 16 12 1780 185 127 39 141 0 16 37 1808 188 129 40 143 2 16 63 18...

Страница 28: ...24 Thermo Scienti c Orion Star A112 Benchtop and Star A122 Portable Conductivity Meters Notes...

Страница 29: ......

Страница 30: ...033 2463887 info water uk thermo com India Tel 91 22 4157 8800 wai asia thermo sher com Japan Tel 81 045 453 9175 wai asia thermo sher com China Tel 86 21 68654588 wai asi thermo sher com Singapore Te...

Отзывы: