18
Thermo Scientific Orion Star A112 Benchtop and Star A122 Portable Conductivity Meters
Meter Specifications
Meter Operating Conditions
Operating Ambient Temperature
5 to 45 °C
Operating Relative Humidity
5 to 85 %, non-condensing
Storage Temperature
-20 to +60 °C
Storage Relative Humidity
5 to 85 %, non-condensing
Pollution
Degree 2
Overvoltage
Category II
Weight
Portable: 450g
Benchtop: 850g
Size
Portable: 5.9 cm (H) x 10.5 cm (W) x 23.1 cm (D)
Benchtop: 9.3 cm (H) x 18.0 cm (W) x 23.6 cm (D)
Regulatory and Safety
CE, TUV 3-1, FCC Class A
Power Rating
DC Input: 9 VDC 1A
Battery: 4 x AA
Shock and Vibration
Vibration, shipping/handling per ISTA #1A Shock, drop test in packaging per ISTA #1A
Warranty
3 years meter replacement
Universal Power Adapter Operating Conditions
Operating Ambient Temperature
0 to 50 °C
Operating Relative Humidity
0 to 90 %, non-condensing
Storage Temperature
-20 to +75 °C
Storage Relative Humidity
0 to 90 %, non-condensing
Pollution
Degree 2
Overvoltage
Category II
Meter Parameter Specifications
Conductivity
Range
0.1 uS/cm to 200 mS/cm
Resolution
0.01uS Minimum;
3 significant figures minimum
Relative Accuracy
0.5% reading ±1 digit
Reference Temperature
20 °C or 25 °C (default)
Compatible Cell Constants
0.001 to 10.00
Number of Calibration Points
1
TDS
Range
1 to 19,999 ppm
Resolution
4 significant digits
Relative Accuracy
0.5% reading ±1 digit
TDS Factor Range
Linear 0.01 to 10.00
Temperature
Range
-5 to 105°C
Resolution
0.1
Relative Accuracy
±0.1
Offset Calibration
1 point
Memory
50 points
Note:
Specifications subject to change without notice.
Содержание Orion Star A112
Страница 1: ...Thermo Scienti c Orion Star A112 Benchtop and Star A122 Portable Conductivity Meters Reference Guide...
Страница 4: ...This page intentionally left blank...
Страница 6: ...This page intentionally left blank...
Страница 28: ...24 Thermo Scienti c Orion Star A112 Benchtop and Star A122 Portable Conductivity Meters Notes...
Страница 29: ......