Thermo Scientific
Heratherm
3-1
3
Installation
Ambient conditions
Location Requirements
Built-in units of incubators can, heating and drying ovens must be operated with an air
exhaust system and exhaust hose (only original Thermo accessory should be used).
For safety reasons, the installation space should be made of non-combustible materials,
according to DIN 4102.
Use with Air Exhaust Systems
For built-in units, a temperature-resistant and corrosion-proof exhaust hose (only original
Thermo accessory) should be used, which can be connected to the air exhaust port with a draft
interrupter.
If several built-in units in a row are connected to a central air exhaust system (see illustration),
a draft interrupter should be installed.
During installation of built-in units, ensure that the
escaping air will be safely discharged out of the
installation space.
Fig. 1
Fig. 2
Содержание OGS 100
Страница 1: ......
Страница 6: ...Contents iv Heratherm Thermo Scientific...
Страница 46: ...Product Description Tube Access Ports 4 16 Heratherm Thermo Scientific...
Страница 60: ...Start up Connecting the RS 232 Interface 5 14 Heratherm Thermo Scientific...
Страница 76: ...Handling and Control Switching the Oven Off Powering Down 7 14 Heratherm Thermo Scientific...
Страница 78: ...Shut down Shutting the Oven Down 8 2 Heratherm Thermo Scientific...
Страница 82: ...Cleaning and Disinfection Wipe Spray Disinfection 9 4 Heratherm Thermo Scientific...
Страница 88: ...Maintenance Returns for Repair 10 6 Heratherm Thermo Scientific...
Страница 104: ...Spare Parts and Accessories 14 6 Heratherm Thermo Scientific...
Страница 106: ...Device Log 15 2 Heratherm Thermo Scientific...
Страница 108: ...Contact 16 2 Heratherm Thermo Scientific...
Страница 109: ...Contact Thermo Scientific Heratherm 16 3...
Страница 110: ...Contact 16 4 Heratherm Thermo Scientific...