-5-
Section II General Information
Description
The NESLAB System IV Liquid to Liquid Heat Exchanger is designed to
remove heat from water-cooled instruments.
The unit consists of a heat exchanger, recirculation pump, stainless steel
reservoir and a microprocessor temperature controller.
Specifications
Temperature Range
Temperature Stability
Cooling Capacity
1
Pumping Capacity
2
Reservoir Volume
Gallons
Liters
Unit Dimensions
3
(H x W x D)
Inches
Centimeters
+5°C to +40°C
±1.0°C
50 gpm at 60 psi
6.5
24.6
33¾ x 23.0 x 27.0
85.7 x 58.4 x 68.6
1. Cooling capacity based on a 10°C difference between the temperature of the cooling water
supply and a 35 gpm process water flow rate; using a 3 horsepower centrifugal pump and
55 psi tap water pressure.
2. CP 100 pump, customizing options available.
3. Additional dimensions are shown on page 19.
5
10
15
20
25
30
35
120
100
80
60
40
20
HEAT
REMOVAL
(KW)
FACILITY FLOW RATE (GPM)
35 gpm process water flow