- 4 -
Section I Safety
Warnings
Warnings are posted throughout the manual. These warnings are designated
by an exclamation mark inside an equilateral triangle and text highlighted in
bold. Read and follow these important instructions. Failure to observe these
instructions can result in permanent damage to the unit, significant property
damage, or personal injury or death.
The lightning flash with arrow symbol, within an equilateral triangle, is intended
to alert the user to the presence of non-insulated "dangerous voltage" within
the unit's enclosure. The voltage may be of significant magnitude to constitute
a risk of electrical shock.
This label, engraved into the front of the tank lip, indicates the presence of hot
surfaces.
Make sure you read and understand all instructions and safety precautions
listed in this manual before installing or operating your unit. If you have any
questions concerning the operation of your unit or the information in this
manual, contact our Sales Department.
Performance of installation, operation, or maintenance procedures
other than those described in this manual may result in a hazardous
situation and may void the manufacturer's warranty.
Transport the unit with care. The units weigh approximately:
EX-7, 27 pounds (12 kilograms); EX-10, 32 pounds (15 kilograms);
EX-17, 36 pounds (16 kilograms); EX-35, 53 pounds (24 kilograms). Units
should be transported with equipment designed to lift these weights.
Observe all warning labels.
Never remove warning labels.
Never operate damaged or leaking equipment.
Never operate the unit without bath fluid in the bath.
The user is responsible for the fluid used. Never use pure ethylene
glycol as a bath fluid, the flash point of 100% ethylene glycol is 111°C.
100% ethylene glycol may produce flammable vapors that can be
ignited by an open flame or an ignition source. When mixed with
water, ethylene glycol is not flammable. Also, at high temperatures
pure ethylene glycol may produce hazardous vapors.
Always turn off the unit and disconnect the line cord from the power
source before performing any service or maintenance procedures, or
before moving the unit.
Always empty the bath before moving the unit.
Never operate equipment with damaged line cords.
Refer service and repairs to a qualified technician.
Содержание NESLAB EX Series
Страница 11: ... 9 Time to Temperature 115V 60 Hz Models 230V 50 Hz Models 230V 50HzModels General Information ...
Страница 12: ... 10 BlankPage ...
Страница 36: ... 34 BlankPage ...
Страница 46: ...BlankPage 42 ...