Preventative maintenance procedures
30
Thermo Scientific™ Multidrop™ Pico 8 Digital Dispense Operating Manual
Thermo Fisher Scientific
Pull the pogo block down so that the pogo pins extend slightly below
the deck.
Inspect the pogo pins from the front and side to ensure that all 48 pogo
pins are not bent and extend fully.
Very gently, in order to not to bend the pogo pins, wipe the pogo pins
with a non-fibrous cloth, such as a Texwipe, saturated with isopropyl
alcohol.
After wiping the pogo pins, push the pogo block back into the
instrument. Allow the isopropyl alcohol to dry prior to next use.
Pogo pin
cleaning