Preventative maintenance procedures
Thermo Fisher Scientific
Thermo Scientific™ Multidrop™ Pico 8 Digital Dispense Operating Manual
29
•
Do not wipe the pogo pins with cleaning fluid other
than isopropyl alcohol.
Clean the pogo pins (see Pogo pin cleaning).
For both the pogo pin inspection and cleaning, turn off the Multidrop
Pico dispenser:
•
Close the Multidrop Pico dispenser software, either by clicking
the X in the upper right corner of the main operating window
or by clicking Exit in the File Menu.
•
Turn off the Multidrop Pico dispenser.
Remove any unused cassettes from the Multidrop Pico dispenser to
avoid contamination and place them back in the cassette tray.
Manually move the well plate to the back-right corner.
Using gloves, gently extend the pogo block out of the instrument and
gently pull it down.
Pogo pin
inspection