Thermo Scientific Medifuge Centrifuge
54
Error number
Description
Troubleshooting
E-40
Acceleration is too low
Make sure that the rotor is correctly loaded
and balanced�
Make sure that no objects are within the
rotor chamber�
Make sure that the correct the mains
connection is correct�
Restart the centrifuge�
If the error message shows again, contact
the customer service�
3. When to contact Customer Service
If you need to contact the customer service, please provide the order no� and the serial no� of your
centrifuge� This information can be found on the nameplate�
In addition the customer service also needs the software ID and the NVRAM ID� Both are shown by
holding the STOP key pressed when switching on the centrifuge�