Operation
Application Screens
3.
Enter a calibration level value. The units displayed are those associated with the
primary level measurement in cal level.
4.
The following commands are available for the detector calibration process:
•
Start CAL – Begin the calibration cycle.
•
Abort CAL – Abort or reject the calibration cycle before cycle completion.
When the user terminates the calibration cycle, the value of the calibration
counts and the level stored in the detector remain unchanged.
•
Accept CAL – Accept the calibration value once the cycle completes. When
the user accepts the calibration value, the values of the calibration counts and
the density at the time of acceptance are stored in the detector.
•
Reject CAL – Reject the calibration value after completion of the calibration
cycle. If the user rejects the calibration, the value of the calibration counts and
the level stored in the detector are unchanged.
5.
The fields listed below are for display purposes only and are not configurable by
the user.
•
Time Remaining (sec)
•
Detector Average Count
•
Calibration/Reference Latest
•
Calibration Temperature
Point Data Tab
(Breakpoint Table)
The Point Data tab displays the minimum and maximum of the level measurement
span, the level of each calibrated point, as well as the point’s count rate and
calibration/standardization ratio.
Figure 4-29.
Application Screen, Gauge Calibration Point Data Tab (Breakpoint Table)
4
-
48
LevelPRO User Manual
Thermo Scientific