3
Connecting the MALDI LTQ XL System
Connecting the MALDI LTQ XL System Forepumps to Line Power
14
LTQ Series Getting Connected Guide
Thermo Scientific
Connecting the MALDI LTQ XL System Forepumps to Line Power
The MALDI LTQ XL system uses two forepumps.The primary forepump provides vacuum to
the LTQ XL MS detector. The secondary forepump provides vacuum to the MALDI sample
module.
As
shows, you connect the primary forepump to the power entry panel of the
LTQ XL MS detector and the secondary forepump to the MALDI control module.
Figure 6.
Line power connections for the forepumps
V
acuum V
alve
Motors
Laser Comm
Sensors
Laser
Power
Aux.
DISCONNECT
WARNING
MODEL:MALDI Laser Control Module
SERIAL No:
CAUTION
Finnigan MAT
PASS
Ground Continuity
and HPOT Tests
Date:
By:
Power
Power In
From LTQ Accessory Outlet ONLY!
OFF
ON
Vacuum Pump
I
0
I
0
MALDI LTQ XL system
(right-side panel)
MALDI control module
(rear panel)
Primary forepump
Power switch
Power switch
Secondary forepump
Power cable
Power cable