ISQ EC and ISQ EM Mass Spectrometer Preinstallation Requirements Guide
Page 23
WARNING—Biohazard
Biohazardous material, for example microorganisms, cell cultures, tissues, body
fluids, and other biological agents can transmit infectious diseases. To avoid
infections with these agents:
•
Assume that all biological substances are at least potentially infectious.
•
Wear personal protective equipment as required by the hazard and follow
good laboratory practice.
•
Dispose of biohazardous waste in an environmentally safe manner that is
consistent with local regulations. Follow a regulated, approved waste
disposal program.
WARNING—Self-Ignition of Solvents
Solvents with a self-ignition temperature below 150 °C might ignite when in
contact with a hot surface (for example, due to leakage in the chromatography
system). Avoid the use of these solvents.
WARNING—Hazardous Vapors
Install the LC-MS or IC-MS system in a well-ventilated laboratory. If the mobile
phase or sample includes volatile or flammable solvents, do not allow them to
enter the workspace. If the mobile phase or sample includes volatile or
flammable solvents, avoid open flames and sparks.
•
The exhaust gas may contain hazardous fumes. To avoid an accumulation of
the exhaust gasses, ensure that the exhaust gas is absorbed by a fume hood
or other ventilating device. Maintain a well-ventilated laboratory. Do not
vent directly into the laboratory.
•
Avoid open flames and sparks. Do not operate the detector in the presence
of flammable gases or fumes.