3
Advanced Maintenance
Replacing Fans and Filters
370
ISQ 7000 Hardware Manual
Thermo Scientific
Intake Filters and RF Shields
To replace the intake filters and RF shields
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Open the front door of the instrument.
3. Use a T10 Torxhead screwdriver to remove the two screws holding the intake filters and
RF shields to the interior front panel.
4. Remove and replace the intake filters and RF shields.
5. Reattach the screws holding the intake filters and RF shields to the interior front panel.
6. Restart your system by following the instructions in
Note
If needed, you can order an additional dust filter. See the
ISQ and TSQ GC-MS
Spare Parts Guide
for ordering information.
Note
You do not have to remove the ion source cartridge or the column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
Figure 421.
Replacing the Filters on the ISQ 7000 Instrument
Intake
Filter/RF
Shield
Screws