Advanced Maintenance
Replacing the Calibration Gas Components
354
ISQ 7000 Hardware Manual
Thermo Scientific
Calibration Gas Module
The EI calibration gas module requires little maintenance, except to refill the calibrant in the
reservoir. Your calibration gas module may need to be replaced if you add too much calibrant
and you contaminate the system. The process is the same for replacing a single- or
dual-calibration gas module.
To replace the EI calibration gas module
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel and left side panel screws
from the interior front panel. Then slide the left and top cover panels toward the back of
the instrument and lift them off.
3. Use a T20 Torxhead screwdriver to remove the support bracket on the left side of the
instrument.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge or the column. However, you
must vent the instrument.
Figure 397.
Removing the Controller Interface Board Support Bracket
Support Bracket
Controller
Interface Board