11. Look through the top of the instrument and use the heat shield to pry the source interface
board off of the 20-pin feedthrough.
12. From the left side of the instrument, use a T20 Torxhead screwdriver to remove the three
screws around the 20-pin feedthrough.
Figure 338.
Detaching the Source Interface Board From the 20-Pin Feedthrough
Figure 339.
Removing the 20-Pin Feedthrough
20-Pin
Feedthrough
20-Pin Feedthrough
Pins in Source
Interface Board
20-Pin Feedthrough
Screws