Advanced Maintenance
Replacing the Covers of the ISQ 7000 Mass Spectrometer
Thermo Scientific
ISQ 7000 Hardware Manual
277
Left Side Panel
To replace the left side panel of the ISQ 7000 instrument
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to loosen the left side panel screw from the interior front
panel.
3. Close the front door.
4. Slide the left side panel toward the back of the instrument and lift it off.
5. Replace the left side panel and tighten the screw holding it in place.
6. Restart your system by following the instructions in
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge or the column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
Figure 291.
Replacing the Left Side Panel