3
Advanced Maintenance
Replacing Components of the Electron Multiplier
Thermo Scientific
ISQ 7000 Hardware Manual
263
Large Feedthroughs
To replace the large feedthroughs, o-rings, and screws
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel screw from the interior
front panel. Then slide the top cover panel toward the back of the instrument and lift it
off.
3. Disconnect the large feedthrough cables.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering
these components.
Note
You do not have to remove the ion source cartridge or the column. However, you
must vent the instrument.
Figure 278.
Disconnecting the Large Feedthrough Cables
Large