3
Advanced Maintenance
Replacing Components of the Ion Source
194
ISQ 7000 Hardware Manual
Thermo Scientific
Source-Repeller Thumbscrews
To replace the source-repeller thumbscrews
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
3. Twist off each source-repeller thumbscrew one at a time to prevent the repeller plate from
falling out.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Figure 177.
Opening the Manifold Door
Figure 178.
Replacing the Source-Repeller Thumbscrews
Note
When you put the spacers back on, make sure you fit them into the grooves on the
thumbscrews. Be sure not to over-tighten the thumbscrews or you may crush the spacers.
Screws
Source-Repeller
Thumbscrews