Repair and maintenance
The PowerEase
™
Touch 600W Power Supply requires no periodic maintenance other than an occasional
dry wipe-down of the instrument.
Encountering Problems
Check the troubleshooting section, see Appendix A, “Troubleshooting”.
Contact Technical Support, see Appendix F, “Documentation and support”
If the unit must be shipped back for repair, contact support for a Return Authorization Number and
shipping instructions; see Appendix F, “Documentation and support”.
Replace the fuse
One extra fuse is supplied with the PowerEase
™
Touch 600W Power Supply. For additional fuses,
contact support; see Appendix F, “Documentation and support”.
1.
Turn off the power switch and detach the power cord at the rear of the power supply unit.
2.
Open the fuse compartment located inside the power inlet by inserting a small flat blade
screwdriver into the slot on top of the fuse compartment. Turn the screwdriver to gently pry open
the fuse compartment.
Note:
The fuse compartment will not open with the power cord in place.
3.
Pull the fuse holder out of the compartment and inspect the fuse. If the fuse is burned or there
is a break in the fuse element, replace the fuse with an identical type of fuse (T10AH/250 V) as
provided in the fuse holder (see Figure 16).
4.
Place the fuse holder back into the compartment.
B
28
PowerEase
™
Touch 600W Power Supply User Guide