Safety
Electrical Safety Precautions
4-8
iCAP Q Operating Manual (P/N 1288090, Revision B)
Thermo Fisher Scientific
Electrical Safety Precautions
WARNING
!
High Voltage.
High voltages capable of causing an electric shock are
used in the instrument.
Observe the following safety precautions when you operate or perform
service on your instrument:
•
Do not change the external or internal grounding connections.
Tampering with or disconnecting these connections could endanger
you and/or damage the system.
•
The instrument is properly grounded in accordance with regulations
when shipped. You do not need to make any changes to the
electrical connections or to the instrument’s chassis to ensure safe
operation.
•
Never run the system without the housing on. Permanent damage
can occur. When leaving the system, make sure that all protective
covers and doors are properly connected and closed, and that heated
areas are separated and marked to protect unqualified personnel.
•
Do not turn the instrument on if you suspect that it has incurred
any kind of electrical damage. Instead, disconnect the power cords
of
mass spectrometer and forepump
and contact a Thermo Fisher
Scientific field service engineer for a product evaluation. Do not
attempt to use the instrument until it has been evaluated. Electrical
damage may have occurred if the system shows visible signs of
damage, or has been transported under severe stress.
•
Do not place any objects upon the instrument—especially not
containers with liquids—unless it is requested by the user
documentation. Leaking liquids might get into contact with
electronic components and cause a short circuit.