Operate the iBright
™
imager
Choose imaging mode and start image capture
On the
Welcome
screen:
1.
Use the drop-down menu to select the desired mode.
•
Chemi Blots
to image chemiluminescent substrates.
•
Nucleic Acid Gels
to image DNA and RNA gels.
•
Protein Gels
to image visible, stained protein gels.
•
Universal
to image samples containing multiple readouts, such as
chemiluminescence, colorimetric stains, and fluorescence (limited to green
transilluminator excitation 490-520 nm and emission 568-617 nm).
2.
Open drawer and place sample in the center of the viewing area.
3.
Close drawer. Touch Live View to see a live-view sample image.
Instrument automatically aligns and focuses on the sample.
Note:
If any manual adjustments are required for zoom or focus, touch
More
options
4
Camera
and choose the appropriate menu item.
• To image in Chemi Blots mode, proceed to “Image using Chemi Blots” on
• To image in Nucleic Acid Gels mode, proceed to “Image using Nucleic Acid Gel”
• To image in Protein Gels mode, proceed to “Image using Protein Gels” on
• To image in Universal Mode, proceed to “Universal Mode” on page 16.
12
iBright
™
CL750 Imaging System User Guide