Universal Mode
Mode to image samples containing one or more signal types (chemi, fluor, and/or
visible).
Table 3 Illumination source, excitation filter, and emission filter for the
iBright
™
CL750 Instrument.
Illumination
Excitation Filters
Emission Filters
Green trans
490-520 nm
568-617 nm
1.
Touch
Choose Channel
and select from one to four different dyes or stains from
the pre-populated list.
a.
To edit an existing dye, touch the dye name button. Touch
Dye
and change
the assigned dye to any other pre-populated dye in the fluorescent dye
list. Dyes and stains in the pre-populated dye list are categorized by most
recently used, channel and name.
b.
Touch
False Color
and change the false color assigned to that dye.
Note:
Each dye is assigned a false color to be used in the composite
overlay.
c.
Touch
Exposure Time
and manually enter the desired exposure time.
Note:
Smart Exposure
d.
To remove this assigned dye from the channel, touch
Remove Dye
at the
bottom left hand corner of the screen and then
Confirm
.
Channel refers to the type of signal you want to capture and the imaging
conditions you want to use.
•
Chemi
: Channel for measuring chemiluminescent signals.
•
Fluor [Trans]
: Channel for measuring fluorescent signals that are
excited by the green transilluminator and appropriate emission filter.
•
Nucleic acid
: Channel for measuring fluorescent signals in gel that are
excited by the green transilluminator and the 568-617 emission filter
(stained DNA and RNA).
•
Protein
: Channel for measuring fluorescent and visible signals in gel
that are excited or illuminated by the green transilluminator and the
568-617 emission filter — stained or labeled protein.
•
Visible
: Channel for measuring visible signals on an opaque media
(nitrocellulose membrane, PVDF membrane, etc.) using the Epi-white
LED with neutral density filter and appropriate emission filter.
•
TPN
: Total protein normalization channel is for normalizing Western
blot results to total protein load by lane.
Operate the iBright
™
imager
Universal Mode
16
iBright
™