Thermo Scientific

S.U.B. TK Resistive Quick Start Guide

Quick Start Guide

 

NOTE:

 Before connecting cables, route them through the tubing and 

cable organizer first
•  If you intend to use the Finesse TruFluor single-use sensors, route 

the appropriate cables from the Tower into the cable management 

arm and down to the sensor window area on the S.U.B. These 

cables will already be attached to the rear of the Control Tower 

via a liquid tight connection and have the following designation, 

DO-2 and pH-2 (No color code).

•  If you intend to use traditional re-usable sensing probes, route the 

appropriate cables in the same manner as described above for the 

single-use sensors. Ensure they are plugged into the back of the 

control tower (

● 

DO-1 

● 

pH-1 

).

•  Locate the bag pressure sensor cable on the Tower (

 ● ●

) and 

route it via the cable management arm and up toward the top arm 

of the tank.

If using a vent heater make the following connections. 

•  Connect cable to labeled VH-1 POWER (

● ●

) power source on 

Tower.

•  Connect Yellow Thermocouple to labeled connection on Tower 

VH-1 FDBK (

● ●

)

•  Route cable through cable management arm and up to vent heater 

brackets on top of S.U.B. unit.

•  If using a second vent heater, use VH-2 connections  

VH-2 POWER (

● ● ●

) and VH-2 FDBK (

● ● ●

) and route 

parallel with VH-1 cable.

Vent Filter Heater

If using a resistive heater make the following connections. 

•  Connect cable labeled RES HEATER (

● ● 

) to the power 

source on the Tower

•  Connect cable labeled HEATER POWER IN (

● ●

 

) into 

designated tower connector.

Resistive Heater 

Power Connection

Содержание HyPerforma S.U.B. TK

Страница 1: ...Thermo Scientific HyPerforma S U B TK Resistive Quick Start Guide...

Страница 2: ...Thermo Scientific HyPerforma S U B TK Resistive Quick Start Guide Rev 1 0...

Страница 3: ...ensuring availability of needed utilities power gas etc Right or left hand orientation was determined at time of order Ensure that tower is located on the same side as the cable and tubing organizer b...

Страница 4: ...wer i Overlay Sparge Black ii Micro Sparge Red iii Macro Sparge White Service gas pressure should be a minimum of 30 psi and a maximum of 35 psi for proper operation Verify that gas connections are se...

Страница 5: ...B TK System Position and attach various hardware items Secure the tool bracket to the tank This position is not critical but would be convenient near the motor Place the ratchet and the custom open en...

Страница 6: ...e use sensors Ensure they are plugged into the back of the control tower DO 1 pH 1 Locate the bag pressure sensor cable on the Tower and route it via the cable management arm and up toward the top arm...

Страница 7: ...ctor must be put in place of the UPS cable 1 3 Power Connections UPS Jumper Figure 5 UPS Jumper Ensure the S U B unit is in its intended position Lock the vessel in place by lowering the leveling feet...

Страница 8: ...front of the tower is not activated should be in the out position If it is engaged then twist to disengage Note that the E stop light will remain red even though it is disengaged until the next step h...

Страница 9: ...he bottom of the BPC Don t remove the protective packaging on top of the BPC at this time 3 Carefully lower the bottom end of the BPC into the S U B with any tubing sets on the top side laying over th...

Страница 10: ...the sample line through the snap in slots on the drain coil arm Attach BPC positioning tabs bottom of chamber onto the alignment pins located at the base of the S U B tank Note that the 50 L BPC does...

Страница 11: ...hrough the main beam of the cable and tubing organizer using the black snap in features Connect to the correct gas fitting labeled MICRO SPARGE Unclamp this lineset when ready to sparge through frit 9...

Страница 12: ...or liquid 13 Position and close a bar clamp Part No SV20664 01 on the bottom drain line as close as possible to the BPC port Figure 11 Also place bar clamps on probe ports intended for conventional p...

Страница 13: ...which is admin To type information without plugging a USB keyboard into the control tower click on the miniature keyboard icon located at the bottom middle of the screen This will bring up a touch scr...

Страница 14: ...uration File button when first turning on the system Click the tab on the top that reads File Select the appropriate S U B TK system to be used by clicking to highlight it Thermo Turnkey 250L Electric...

Страница 15: ...Thermo Scientific 13 S U B TK Resistive Quick Start Guide Quick Start Guide Figure 17 Home Monitor View...

Страница 16: ...Guide Quick Start Guide 2012 Thermo Fisher Scientific Inc All rights reserved Cell Culture BioProcessing 925 West 1800 South Logan UT 84321 InAmericas Asia 435 792 8000 435 792 8001fax InEurope 32538...

Отзывы: