7. Route the overlay sparge set (located at the top of the BPC marked
“Overlay Sparge”) through the cable and tubing organizer arm to
the top of the main beam of the cable and tubing organizer and
connect to the fitting marked OVERLAY SPARGE. This will be
the lineset that will be used to pre-fill the BPC with air, so ensure
that the clamp on this lineset is released.
8. Route the Micro Sparge Line (labeled), located at the bottom
of the BPC, up through the main beam of the cable and tubing
organizer using the black snap in features. Connect to the correct
gas fitting labeled MICRO SPARGE. Unclamp this lineset when
ready to sparge through frit.
9. Route Macro Sparge line (labeled), located at the bottom of
the BPC, running parallel to the Micro Sparge Line, using the
same snap in features. Connect to the correct gas fitting labeled
MACRO SPARGE. Ensure the clamp is released on this line, as
well before operation.
NOTE:
To reduce the risk of filter membrane wetting, position
each sparge filter above the low point of the sparge line.
10. Connect the pressure sensor cable onto the sensor located at the
top of the BPC.
Figure 10.
Pressure sensor
Figure 9.
BPC linesets