HistoStar Operator Guide A81010100 Issue 2
84
Base Mould Storage
Problem
Possible Cause
Remedy
Base mould storage does
not heat up
Base mould storage
heater is not switched
on
Ensure heater is switched on - see Temperature Set Points
Ensure temperature setting is correct
Compare the set point temperature to the actual temperature -
if the set point temperature is not achieved after a suitable
length of time:
•
Switch off the embedding module
•
Wait for 20 minutes and then switch the embedding
module back on
If problem persists call Thermo Fisher Scientific Service
Department
Instrument
malfunction
Switch the embedding module off for 30 seconds and then
back on again to reset the control system
Perform Heater System Check
Call Thermo Fisher Scientific Service Department