Maintenance and Care
Disinfection
5-4
Multifuge 4KR
Thermo Scientific
Disinfection
Disinfect the centrifuge immediately whenever infectious material has spilled during centrifugation.
The rotor chamber should be treated preferably with a neutral disinfectant.
Contact the Service Department of Thermo Fisher Scientific for questions regarding the use of other
disinfectants.
Disinfect the rotor and accessories as follows:
1. Open the centrifuge.
2. Turn off the centrifuge.
3. Pull out the power supply plug.
4. Release the rotor.
5. Grasp the rotor with both hands and lift it vertically off the centrifuge spindle.
6. Treat the centrifuge according to the instructions for the disinfectant. Adhere strictly to the given
application times.
7. Wipe off the disinfectant.
8. Rinse thoroughly with water.
9. Dispose of the disinfectant according to the applicable guidelines.
Decontamination
Decontaminate the centrifuge immediately whenever contaminating material has spilled during
centrifugation.
WARNING
Infectious material can get into the centrifuge when a tube breaks or as a
result of spills. Keep in mind the risk of infection when touching the rotor and take all
necessary precautions.
In case of contamination, make sure that others are not put at risk.
Decontaminate the affected parts immediately.
Take other precautions if need be.
CAUTION
Before using any cleaning or decontamination methods except those
recommended by the manufacturer, users should check with the manufacturer that the
proposed method will not damage the equipment.
Observe the safety precautions and handling instructions for the cleaning agents used.
WARNING
Infectious material can get into the centrifuge when a tube breaks or as a
result of spills. Keep in mind the risk of infection when touching the rotor and take all
necessary precautions.
In case of contamination, make sure that others are not put at risk.
Decontaminate the affected parts immediately.
Take other precautions if need be.