Mechanical
FT6060
Unit
Dimensions
Housing (W x H x D (incl. door handle)
600 x 794 x 640
mm
Inner chamber (W x H x D)
403 x 380 x 339
mm
Trays (W/D)
387 / 335
mm
Chamber vol.
approx. 52
l
Weights
Unit (empty)
approx. 92
kg
Maximum total load
30
kg
Max. inserted load (distributed)
15
kg
Connections
Fresh air (inside pipe
∅
)
100
mm
Exhaust air (inside pipe
∅
)
100
mm
Fan fail-safety device
input pressure setting: 1 bar
(inert gas or compressed air)
∅
5, for hose
∅
4 x 1
mm
Electrical
FT6060
Unit
Nominal voltage, nominal frequency
1 / PE AC, 230 V; 50 Hz
Connected load
1.46
kVA
Power consumption
1.32
kW
Protection class
I
Degree of
protection
Electrical equipment
IP 54
Inner chamber
explosion-protected design
Circuit protection
(Only use properly installed socket-outlets (with PE conductor) to connect the unit to
the mains supply. Comply with the applicable national electrical engineering
regulations and technical requirements)
T 16 A fusible link or B 16
circuit-breaker,
Connection via an earth-leakage circuit breaker
(tripping current < 30 mA)
Thermal
FT6060
Unit
Operating temperatures up to
105
°C
Spatial temperature deviations with an
operating temperature of
105°C
1)
+ 3.0
°C
70°C
1)
+ 2.0
°C
Temporal temperature deviations at 105 °C
1)
< 0.5
K
Warm-up time, from 25 °C to approx. 98 %
of the operating temperature:
105 °C
1)
50
Min.
70°C
1)
50
Min.
Heat radiation to the surrounding areas
1.3
kW
1)
Measured in accordance with DIN 12 880 Part 2
Noise (in accordance with DIN 45635)
FT6060
Unit
Sound level
57
dB(A)
Sound power level
71
dB(A)
6.
TECHNICAL DATA
50 043 136
23/25