Chapter 15
Maintenance and Troubleshooting
Maintaining a FID
286
Figure 15-13. Clean/Replace the Jet (2)
•
then continue from the point 7.
•
then continue from the point 7.
Clean the Jet
a.
Place the jet in the ultrasonic bath filled with aqueous detergent and clean
it for about five minutes.
b. Handling the jet with forceps or tweezers, rinse the jet with distilled
water, then with methanol.
c.
Place the jet on a paper towel and let it air dry. When the jet is dry, place
the jet into the housing and tighten it with the proper tool. Ensure the jet is
perfectly vertically aligned to avoid damaging its ceramic part.
Replace the Jet
a.
Place a new jet into the housing and tighten it with the proper tool. Ensure
the jet is perfectly vertically aligned to avoid damaging its ceramic part.
See Figure 15-14.
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...