Chapter 15
Maintaining a FID
Maintenance and Troubleshooting
287
Figure 15-14. Clean/Replace the Jet (3)
7. Place the detector cell on the detector base body and tighten the fixing screw
on front of the detector cell.
Reconnect the signal and ignition cables to the detector. See Figure 15-15
Figure 15-15. Clean/Replace the Jet (4)
8. Reset the detector to the required operating conditions.