Isotemp Freezer
Installation | 7
Installation
Electrical
Units must be connected to a grounded outlet matching the
name plate and / or the information furnished in this manual. If
you are not sure about the outlet, you should contact a
qualified electrician for assistance.
Be Advised
Selecting a Location
Choose a location for the unit that will provide a clearance of
3/8 in. at the top, 1 in. at the rear and 3/8 in. at each side.
Appropriate electrical power must be available. Locate the
unit near to the power outlet so that no extension cord is
required. Attach the unit to the facility’s electrical supply as
directed by the National Electrical Code.
Leveling the Unit
This unit must be leveled in order to provide adequate
condensate drainage as well as proper door alignment and
operation. This unit should be in its final operating location
and set, so that it is firmly positioned on the floor.
Setting the anti-tip bracket:
Unit is equipped with an anti-tip bracket. The anti-tip brackets
are located on the lower front corners of the cabinet. Lower
the anti-tip on each side clockwise until they contact the floor.
Do not raise the cabinet.
Figure 1. Anti-tip bracket
Note:
For slippery/tiled floor locations, it is suggested to place
locally purchased anti-skid pads under the leveling legs to
prevent chances of sliding.
WARNING:
DO NOT REMOVE, under any
circumstance, the grounding prongs from the
3-prong power cord supplied with all units.
WARNING:
DO NOT USE electrical extension
cords, it may result in voltage loss and possible
hazardous operation.
CAUTION:
Do not block the lower front of your
unit. Sufficient air circulation is essential for the
proper operation of your unit.
WARNING:
Storage by user of any materials in
the product that may cause a deterioration of the
product shall be deemed to constitute abnormal
and improper usage of the product for purposes
of this warranty.
WARNING:
Before connecting the final power
supply, check the electrical characteristics of the
unit nameplate to see that it is in agreement with
the power supplied. In addition, power should be
wired to the unit according to the electrical
schematic and all applicable codes. Only
qualified electricians should work on the
electrical portion of any unit installation.
WARNING:
The anti-tip device must be installed
according to the instructions in your User
Manual. Failure to do so will result in injury.
Lower
Содержание Fisher brand Isotemp FBV20FPSA
Страница 1: ...Isotemp Freezer Installation and Operation 331487H03 Revision A 7 15 2020...
Страница 26: ...Cong lateur Isotemp Installation et fonctionnement 331487H03 R vision A 7 15 2020...
Страница 51: ...Cette page est intentionnelleent laiss vide...
Страница 52: ...Cette page est intentionnelleent laiss vide...