background image

 

                                                                                                  

42 

           

 Ordering Procedures    

 
 

 
 
 
 
 

 

Please refer to the Specification Plate for complete model 
number, serial number, and series number when request- 
ing service, replacement parts or in any correspondence 
concerning this unit. 

 

All parts listed herein may be ordered from the Thermo Scientific 

dealer from whom you purchased this unit or can be obtained 
promptly from the factory. When service or replacement parts are 
needed, we ask that you check first with your dealer. If the dealer 
cannot handle your request, then contact our Customer 
Service Department at 800-438-4851. 

 

Prior to returning any materials to Thermo 

Scientific

please contact our Customer Service Department for a 
“Return Materials Authorization” number (RMA). 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание FD1500M

Страница 1: ...1 Installation and Operation Manual Thermo Scientific Thermolyne Furnace Benchtop Industrial Type FD1500M Type Models FD1530M FD1530MCN FD1530M 33 FD1535M FD1540M FD1540MCN FD1540M 33 AND FD1545M...

Страница 2: ...written consent of Thermo Fisher Scientific Trademarks Thermo Scientific is a brand owned by Thermo Fisher Scientific Inc All trademarks mentioned in the operating instructions are the exclusive prop...

Страница 3: ...ower Switch 11 Cycle Light 11 Door Safety Switches 11 Digital Readout 11 Single Setpoint Model w OTP 12 Eurotherm 3216 Controller Operation 12 Basic Operation 12 Button and Indicators 12 To View or Ch...

Страница 4: ...8 Segment Programmable Models 29 Holding a Program 29 Cancelling a Program 29 Tuning 29 Autotuning 31 Adaptive Tuning 31 Furnace Loading 32 Preventative Maintenance 33 Troubleshooting 34 Maintenance...

Страница 5: ...liability and safety in mind It is the user s responsibility to install it in conformance with local codes For safe operation please pay attention to the alert signals throughout the manual Warnings T...

Страница 6: ...Group 1 The insulating materials can be located in the door the hearth collar in the chamber of the product or under the hot plate top Tests performed by the manufacturer indicate that there is no ri...

Страница 7: ...he furnace for more than 3 hours and intermittent use is operating the furnace for less than 3 hours The unit consists of 1 heating chamber 2 a digital controller and 3 a door interlock relay for user...

Страница 8: ...M OVERALL DIMENSIONS IN CM WIDTH HEIGHT 11 28 CM 16 1 2 42 CM 11 28 CM 16 1 2 42 CM 11 28 CM 16 1 2 42 CM DEPTH 18 46 CM 18 46 CM 18 46 CM WEIGHT LBS KG 42 LBS 19 KG 42 LBS 19 KG 42 LBS 19 KG ELECTRIC...

Страница 9: ...ith IEC 664 Pollution Degree 2 in accordance with IEC 664 Altitude limit 2 000 meters Storage 25 C to 65 C 20 to 80 relative humidity Declaration of Conformity For 33 models only Thermo Scientific her...

Страница 10: ...The owner s manual and door handle are included in the box After unpacking the furnace attach the door handle and remove packing material from inside furnace chamber The furnace is supplied with one...

Страница 11: ...ermo Scientific if your electrical service is different than those listed on the specification plate Prior to connecting your Type 1500 furnace to your electrical supply be sure the two position power...

Страница 12: ...hed ON The furnace will begin to heat to the controller s current setpoint See the instructions for your type of controller for information on checking and setting the setpoint Cycle Light The amber c...

Страница 13: ...upper display and the setpoint is found in the lower display Buttons and Indicators OP1 Output 1 Illuminates when the output is ON normally heating OP2 Output 2 Illuminates when the output is ON norm...

Страница 14: ...he HOME DISPLAY to the right of the measured value process value Temperature units can be changed by pressing up and down buttons Choice of Celsius C Fahrenheit F Kelvin K Percentage or None nonE Cont...

Страница 15: ...roperly Ld F Heater Circuit fault indication of either an open or short solid state relay a blown fuse missing supply or open circuit heater Sensor Break Protection This controller provides sensor bre...

Страница 16: ...rt tuning with the process at ambient temperature This allows the tuner to calculate the low cutback and high cutback values more accurately 2 Adjust the setpoint to your desired value 3 Press the SCR...

Страница 17: ...L The dwell time begins once the setpoint reaches the set threshold The END TYPE action is executed when the dwell timer reaches the end 2 Mode 2 DELY The timer starts immediately upon instrument powe...

Страница 18: ...amp Rate is displayed 2 Set the desired Ramp rate with the UP or DOWN buttons if the ramp to setpoint feature is needed If the Ramp rate is not needed then set to OFF with the feature is needed If the...

Страница 19: ...Actual temperature Stopping the Program Press the SCROLL button until T STAT is displayed Set to res with the UP or DOWN buttons Clearing the Flashing End Press the PAGE and SCROLL buttons at the sam...

Страница 20: ...ller The 8 segment digital model enables eight segments of programming Basic Operation When the controller is turned ON it will perform a short self test and then change to the HOME DISPLAY The HOME D...

Страница 21: ...n Starts a program when pressed once RUN light illuminates Holds a program when pressed again HOLD light illuminates Cancels hold and continues running when pressed again HOLD light is off and RUN lig...

Страница 22: ...e lower display The controller will default to OFF FASt Fast run through program no YES The controller will default to no SEG d Flash active segment type in the lower display of the home display no YE...

Страница 23: ...default to End Other parame ters used with tYPE include tGt target setpoint Rate rate of temperature increase and dur time to target setpoint or time to dwell End t End segment type dwELL dwell contin...

Страница 24: ...ly L br Loop Break Check that the heating circuits are working properly Ld F Heater Circuit Fault indication of either an open or short solid state relay a blown fuse missing supply or open circuit he...

Страница 25: ...oint Controller 1 Switch the circuit breaker to the ON position The setpoint temperature presently set in the controller will appear in the lower display The upper display indicates the actual chamber...

Страница 26: ...he program back when process variable deviates above the setpoint by more than the holdback value bAnd Deviation Band Holdback combines the features of the high and low deviation holdback in that it h...

Страница 27: ...are used in dwELL segments amount of time to remain at a specific temperature See Setting the Segment Type for an explanation on how to set a dwell segment 1 Press the PAGE button until you reach the...

Страница 28: ...s designed to reduce the heatup rate or cooling rate that the furnace normally exhibits When not using this feature the furnace will operate at its maximum heating and cooling capability Note When the...

Страница 29: ...GE button until you reach the pro gram list ProG LiSt 2 Press the SCROLL button until display reads tYPE 3 Press the UP or DOWN button until display reads dwEll 4 Press the SCROLL button until display...

Страница 30: ...l an indefinite dwell at last target level S OP End Segment Output Power and rSET reset will revert back to setpoint in Single Setpoint mode Running a Program 8 Segment Programmable Models To run a pr...

Страница 31: ...edure Tuning Error The display will flash tu ER if an error occurs during tuning To clear the error and restart tuning simultaneously press the PAGE and SCROLL buttons and follow the steps outlined in...

Страница 32: ...ress the PAGE button until the display reads Atun LiSt 4 Press the SCROLL button until tunE OFF is displayed Note Start tuning with the process at ambient temperature This allows the tuner to calculat...

Страница 33: ...be heated uniformly it should not occupy more than the center two thirds of the furnace chamber Failure to observe this caution could result in damage to furnace components Furnace Loading 1 For best...

Страница 34: ...ge All heating elements must be considered expendable and replacement is expected however reasonable care in their use will greatly extend the service they will give As the manufacturer has no control...

Страница 35: ...p Repair electrical hookup Thermocouple has oxidized and opened the circuit Replace thermocouple Controller malfunction Replace controller Heating elements burned out Replace defective elements Solid...

Страница 36: ...ision of control setting Heating harmful materials Enclose material in container Clean up spills in and on chamber Ventilate chamber by leaving top vent slightly open when heating known harmful reagen...

Страница 37: ...urn the furnace so that the element to be removed is on top 6 Slide the new element into place threading the leads through the insulating porcelain bushing in the back of the furnace 7 Reverse the dis...

Страница 38: ...Replace the back cover 9 Reconnect furnace to power supply 10 Test the operation of the furnace To Replace Door Switch Micro Switch 1 Disconnect the furnace from power supply 2 Remove the screws from...

Страница 39: ...he door approximately 1 and rotate the cam counterclockwise until you hear a click then rotate the cam clock wise until you hear a click tighten the set screw on adjusting cam 3 Open and close the doo...

Страница 40: ...30M FD1530MCN FD1535M CN71X159 Temperature Control Programmable 2 ramp 2 dwell FD1540M FD1540MCN FD1545M CN71X179 Temperature Control Single Set Point FD1530M 33 CN71X160 Temperature Control 8 Segment...

Страница 41: ...40 Wiring Diagrams Single Setpoint w OTP Control...

Страница 42: ...41 WIRING DIAGRAMS...

Страница 43: ...om the Thermo Scientific dealer from whom you purchased this unit or can be obtained promptly from the factory When service or replacement parts are needed we ask that you check first with your dealer...

Страница 44: ...Purchased The above information can be found on the dataplate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or specific agent rep organiz...

Страница 45: ...Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States www thermofisher com LT1170X1 Rev B...

Отзывы: