Using the Spectrophotometer
8
Evolution Pro Spectrophotometer User Guide
Thermo Scientific
Tip
The power light flashes slowly when the spectrophotometer is starting up. When the
start-up sequence is complete the light will be steady. If the power light starts to flash
rapidly this indicates an error condition. Check INSIGHT software for any error reports.
If INSIGHT it not running or shows no error
1.
Power the spectrophotometer OFF.
2.
Perform basic checks and prepare to collect fault information:
a.
Check that both sample and reference beams are clear of obstacles.
b.
Close the sample compartment doors.
c.
Start INSIGHT software on the attached computer.
d.
Power the spectrophotometer ON.
e.
Watch the progress of the startup routine.
3.
If the error repeats, note the stage (bottom left corner of the window) where the error
occurred and contact us.
•
Trigger port – connect a contact closure trigger device here
–
Used with the rapid mix kinetics accessory
–
Used with third party remote triggering devices
•
Accessory Port – provides support for planned future accessories
•
USB-B Port – connect the computer running the instrument control software here
Using the Spectrophotometer
Connecting and Powering the Spectrophotometer
Refer to the Getting Started guide supplied with the instrument (printed copy) and also on
the same media as this user guide.
Making a Measurement
Refer to the INSIGHT software help system, included with your installation of INSIGHT
software. INSIGHT software help includes quick-start guides for each application mode.
These are available with one click from the help system home screen.
The beam direction in the sample compartment is from right to left.
The beam direction in the reference compartment is from front to back.