background image

 Useful Notes

1. To avoid cross contamination, rinse with clean water between samples and calibration standards. 

2. For long-term storage, fill the cap with 20 mm of storage solution (NOT de-ionized water). 

This will keep the sensor conditioned and ready for next use.

3

.  Not recommended

 for prolonged use with solutions containing heavy metals, proteins, tris buffer, or 

sulfides which will clog the junction leading to slow response and eventual failure.

 Changing Batteries (4) A76 or LR44 button cell type

1. Holding the tester with one hand, slide in the thumb to clear the front catch. 
2. While still holding the tester, slide in the other thumb to clear the back catch. 
3. With both catches are cleared, vertically slide the battery cover off the tester.
4. Change the batteries noting the polarity (flat positive side to the left). Replace 

the battery cover onto the tester with the shorter tab above the display locking
the front and back to ensure a watertight seal. 

 Maintenance

1. Rinse sensor with clean water after each use.
2. Clean the sensor with a soft brush and mild dish soap to remove dirt and grease.

 Error Messages

1.

 –Batteries are weak and need replacement soon.

2

. bAt Lo

 (low battery) – The battery power is too low to power on the instrument and requires 

 iimmediate replacement. 

3

. StBL Err

 ( stabilizing error) – Appears when calibration is attempted but the reading is not yet 

 stable. Wait for the reading to stabilize or manually confirm the calibration by pressing enter.

3.  

bUFF Err

 (buffer error) – The buffer is outside of the calibration range.

4.  

SLPE Err

 (slope error) – The 2

nd

 or 3

rd

 calibration point is not within 80% to 120% slope range.

5.  

Or

 (over range) – The reading is above the measuring range of tester.

6.  

Ur

 (under range) – The reading is below the measuring range of tester.

Testers and Accessories Order Information

Ordering Code

Product Description

Pocket Tester

EXPERTPH

pH pocket tester with batteries

pH Buffer Solutions and Sachets

ECBU4BT

Colourless pH 4.01, 480 mL

ECBU7BT

Colourless pH 7.00, 480 mL

ECBU10BT

Colourless pH 10.01, 480 mL

ECPHBUFKIT

Colourless pH buffer set, pH 4.01, 7.00, 10.01, 480 mL each

ECBU4BS

pH 4.01 (NIST traceable), box of 20 x 20 mL sachets

ECBU7BS

pH 7.00 (NIST traceable), box of 20 x 20 mL sachets

ECBU10BS

pH 10.01 (NIST traceable), box of 20 x 20 mL sachets

ECRINWT

pH deionized water (NIST traceable), box of 20 x 20 mL sachets

ECBU4710R

pH buffer sachet set, pH 4.01, 7.00, 10.01, deionized  water, 5 x 20 mL each

Storage and Cleaning Solutions

ECRE005

Storage solution for pH sensor, 480 mL

ECDPCBT

Protein remover solution, 480 mL

Other Accessories

EXPERTCAP

Replacement sensor cap

LANYARD

Tester lanyard

 Warranty

This instrument is supplied with a warranty against manufacturing defects for a period of one year  
from the date of purchase.    

Operating Instructions For Expert pH

  Conditioning Before First Use

The pH sensor must be hydrated to perform properly. If the pH electrode has been stored dry, soak as long as 
needed to re-hydrate the sensor. Electrode storage solution works best, followed by pH buffer solution, followed 
by tap water. 10 minutes is usually adequate but it may take much longer if the bulb is extremely dry.  The 
transparent cap can be used for soaking and storage. Keep upright for best results. Never store the sensor for 
extended periods of time in deionized water.

  

Measurement

1. Remove the cap and press the   to power on the tester.
2.  Dip sensor in at least 20mm of test solution or up to the MAX fill line of the cap.
3.  Stir gently while the reading stabilizes and wait for the    icon to stop blinking. 
4. The   will appear on the display indicating that the reading is stable and measurement is complete.

Note: The tester automatically shuts off after 8.5 minutes of non-use to conserve batteries if you forget to turn

   it off.

 Menu (To Change Settings)

      

Press 

MENU

 to begin, and to leave the menu at anytime. Press 

CAL

 to return to the measurement mode.

To choose USA (4.0, 7.0, 10.0) or NIST (4.0, 6.9, 9.2) buffer values for automatic calibration:

To reset the calibration to factory default condition:

To change celsius / fahrenheit temperature unit:

To adjust the temperature reading:

  Calibration

For best results, periodic calibration with accurate buffers is recommended prior to measurement. Use up to 3  
standards and bracket your intended measuring range in any order. The tester will return to measurement after 
each calibration point and retain the calibration when the instrument is powered off. 

1. Remove the cap and press the   to power on the tester. 
2.  Dip the sensor in at least 20mm of pH buffer.
3.  Stir gently and press 

CAL

  to begin the pH calibration.

4. The display will show 

CAL 

followed by the default pH value.   is indicated on the display during calibration 

mode. 

5. The   icon will stop blinking when the pH reading is stable.
6. The   icon is displayed when the automatic buffer value has been detected. 
7.  After 3 seconds, the tester will automatically calibrate the value. Alternatively, pressing       will also accept 

the value manually. Tester will show 

donE

 to confirm the calibration.

8. To calibrate additional point(s), repeat with additional pH buffers.
9.  To abort pH calibration, press 

CAL

  to escape.

bUFF

bUFF

USA

bUFF

nlSt

donE

bUFF

r.SEt

r.SEt

nO

r.SEt

yES

r.SEt

bUFF

donE

r.SEt

dEG

dEG

 C

°

C

dEG 

  F

°

F

donE

dEG

dEG

t.CAL

78.9

78.9

°

F

78.9 

77.0

°

F

donE

t.CAL

MENU

MENU

MENU

MENU

MENU

MENU

ENTER

MENU

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

Отзывы: