Configuring the Mass Spectrometer for NSI Mode
Configuring the EASY-Spray Source’s NSI Parameters
32
EASY-Spray Series Ion Source User Guide
Thermo Scientific
Figure 20.
Ion Source pane in the Tune window (example)
4. Click
Apply
.
Configuring the EASY-Spray Source’s NSI Parameters
After you complete the instrument configuration, use the Thermo Tune Plus application to
configure the NSI source parameters. For additional information, refer to the Tune Plus Help.
To set the NSI source parameters
1. Open the Tune Plus application by choosing
Start > All Apps
(Windows 10) or
All
Programs
(Windows 7)
> Thermo Instruments
(or
Thermo
Model
) and so on.
2. Choose
Setup > NSI Source
, and then enter
1.90
in the Spray Voltage (kV) box
).
Use 1.90 kV as the start value for the spray voltage. If the intensity of the full-scan
spectrum is low, gradually increase the spray voltage to improve the spectrum. The
recommended range for the spray voltage is 1.40–2.40 kV.
Figure 21.
NSI Source dialog box
3. Click
OK
.
Enter the
absolute value.