Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
76
DSQ II Hardware Manual
Thermo Scientific
5. Remove the calibration gas flow module.
Figure 42.
Removing the Calibration Gas Flow Module
a.
Disconnect the gas fittings.
b. Disconnect the foreline adapter tubing from the foreline adapter, which is located
near the high vacuum pump.
c.
Remove the two screws.
d. Remove the calibration gas flow module.
6. Install the new calibration gas flow module (PN 119246-0001).
a.
Insert the new module.
b. Reattach the two screws.
c.
Reattach the foreline adapter tubing to the foreline adapter.
d. Reconnect the gas fittings.
1
Gas Fittings
4
Two Screws
2
Foreline Adapter Tubing (not
shown)
5
Two Twisted Pair Wires
3
Calibration Compound Vial
6
Calibration Gas Flow Module
(PN 119246-0001)
3
1
2
6
4
5