Troubleshooting
Power Supply Symptoms
Thermo Scientific
DSQ II Hardware Manual
139
Power Supply Symptoms
DSQ II Tune
diagnostics detects most power supply symptoms. Power supply problems often
involve a blown fuse, faulty electronic components, or even something as simple as a
disconnected cable. Use extreme care when you replace a fuse or electronic component.
DSQ II will not turn on.
Possible Solutions
• Disconnected power cord. Verify that the power cord is plugged in.
• Voltage not coming from the electrical outlet. Verify that the electrical outlet is
operational.
• Improper voltage at the electrical outlet. Verify that the outlet is using the correct voltage
for your system.
• Faulty power module. Replace power module, as described in
DSQ II turns on, but trips the circuit breaker.
Possible Solutions
• Faulty power module. Replace power module, as described in
• Rotary-vane pump causes circuit breaker to trip. Check the rotary-vane pump voltage
setting; replace if necessary.
WARNING - ELECTRICAL SHOCK HAZARD.
Before proceeding. unplug the DSQ II to
avoid an electrical shock.
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...