2
Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
100
DSQ II Hardware Manual
Thermo Scientific
6. Remove the low pass filter board.
Figure 59.
Removing the Low Pass Filter Board
a.
Turn the RF module over (setting it down on its top), and remove the two screws
holding the low pass filter cover in place.
b. Unscrew the connectors to the J2, J4, J5, J6, J7 wires.
c.
Disconnect the J1 cable.
d. Disconnect the P1, P2, and P4 cables.
e.
Remove the four screws holding the low pass filter board in place.
f.
Remove the current low pass filter board.
1
Connector to J1 Cable
8
Connector to J6 Wire (gold)
2
Screws Around the Low Pass Filter
Cover
9
Four RF Feedthrough Connections
3
Connector to P1 Cable (green)
10
Connector to J2 Wire (white)
4
Connector to P2 Cable (black)
11
Connector to J5 Wire (red)
5
Connector to P4 Cable (red)
12
Connector to J4 Wire (green)
6
Low Pass Filter Board
(PN 119590-0480)
13
Four Screws Around the Low Pass
Filter Board
7
Connector to J7 Wire (gold)
9
1
2
3
4
5
6
7
8
2
10
11
12
13
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...