2
Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
Thermo Scientific
DSQ II Hardware Manual
97
a.
Disconnect the cables to the low pass filter board.
b. Disconnect the cable to the analog board.
Figure 56.
Removing the Lens Interface Board
c.
Remove the four screws holding the lens interface board
in place.
d. Remove the current lens interface board.
5. Install the new lens interface board (PN 119590-0110).
a.
Insert the new lens interface board.
b. Reattach the screws that hold the lens interface board in place.
c.
Reconnect the cables.
6. Reattach the right, left, top, and front covers of the DSQ II.
7. Restart the DSQ II.
1
Lens Interface Board
(PN 119590-0110)
3
Vacuum Manifold
2
Four Screws Around the Lens
Interface Board
2
1
3