Service Procedures
Doc. 031880-04 1/12
51
6. Lift the EGC out
of the holder (the
inlet and outlet
tubing is
attached). Turn
the EGC upside
down and set it
next to the holder
(see Figure 45).
Figure 45. EGC Inlet and Outlet Connections
7. Disconnect the
EGC IN
and
EGC OUT
lines from the EGC inlet and
outlet ports.
8. Follow the instructions in the EGC manual (Document No. 031908),
on the Thermo Scientific Reference Library DVD (P/N 053891), to
prepare an expended EGC for disposal or to store a partially used
EGC for later use.
9. Turn the new EGC upside down and set it next to the holder.
10. Remove the plugs
from the EGC
INLET
and
OUTLET
fittings
Figure 46. EGC Fitting Plugs
Inlet
Outlet
Fitting Plugs
Outlet
Inlet