Operation
Doc. 031880-04 1/12
17
4. Press
ENTER
to save the value.
Setting the Dionex RFC-10 Current
For optimal suppressor performance, use the suppressor current
recommended for your target application. The setting to use depends on
several variables: the suppressor type, the column, the eluent type and
concentration, and the flow rate. Refer to the suppressor manual for the
recommended current setting for your application. Suppressor manuals
are included on the Thermo Scientific Reference Library DVD
(P/N 053891).
Before setting the current, verify that the correct suppressor type and flow
rate are selected. The maximum current allowed for the suppressor
depends on its type, and for the Atlas suppressors, the flow rate (see
Table 2).
SRS Type
Flow Rate Range
Dionex ASRS_2MM
0.10 – 1.00 mL/min
Dionex ASRS_4MM
0.10 – 3.00 mL/min
Dionex CSRS_2MM
0.10 – 0.75 mL/min
Dionex CSRS_4MM
0.10 – 3.00 mL/min
Dionex AAES
0.10 – 3.00 mL/min
Dionex CAES
0.10 – 3.00 mL/min
Table 1. Flow Rate Ranges per Suppressor Type
Suppressor Type
Current Range
Dionex ASRS_2MM
0 – 150 mA
Dionex ASRS_4MM
0 – 500 mA
Dionex CSRS_2MM
0 – 110 mA
Dionex CSRS_4MM
0 – 300 mA
Dionex AAES
0 – 100
(
Flow
)
0.6
mA
a
a. The maximum current cannot be greater than 150 mA.
Dionex CAES
0 – 100
(
Flow
)
0.6
mA
a
Table 2. Current Range Allowed for Each Suppressor Type