Pressurizing the Eluent Reservoir (Optional)
Doc. 031857-03 4/06
B-115
B.21.2 Pressurize the Eluent Reservoir
NOTE Do not pressurize the ICS-2000 until after the
reservoir is plumbed.
1. Pull out the air regulator knob and turn it fully counterclockwise to
ensure there is no pressure to the reservoir when the gas source is
turned on.
2. Turn on the gas source.
3. Pull out the air regulator knob and turn it clockwise to pressurize the
eluent reservoir.
4. Adjust the pressure to between 0.03 and 0.04 MPa (5 and 6 psi). Push
the regulator knob back in.
Do not pressurize the eluent reservoir above 0.07 MPa (10 psi).
Ne mettez jamais les réservoirs d'éluants sous une pression
supérieure à 0,07 MPa (10 lb/po²).
Setzen Sie den Eluentbehälter auf keinen Fall einem Druck über 0,07
MPa aus.
Содержание Dionex ICS-2000
Страница 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Страница 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Страница 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Страница 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Страница 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Страница 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...