5 • Service
Doc. 031857-03 4/06
5-37
8. Close the waste valve.
Do not overtighten.
9. Select the flow rate required for the analysis.
10. Click
Pump Off
to turn off the pump motor. The pump is now ready for
operation.
5.16 Changing Main Power Fuses
1. Turn off the main power switch.
2. The fuse holder is part of the main power receptacle (see
) on the
ICS-2000 rear panel. To remove the fuse holder, squeeze the tab on the top of
the holder to release it and pull the holder straight out of its compartment.
Isopropyl alcohol may damage some columns. Be sure to thoroughly
rinse the alcohol from the pump, using the process described in
HIGH VOLTAGE—Disconnect the main power cord from its source and
also from the rear panel of the ICS-2000.
HAUTE TENSION—Débranchez le cordon d'alimentation principal de
sa source et du panneau arrière du ICS-2000.
HOCHSPANNUNG—Ziehen Sie das Netzkabel aus der Steckdose und
der Netzbuchse auf der Rückseite des ICS-2000.
Figure 5-30. Main Power Fuse Holder