4 • Troubleshooting
Doc. 155028-01 08/20
55
•
Flow rate through columns too high
Check the pump flow rate.
•
Clogged column bed supports
Refer to the column manual for troubleshooting procedures.
•
Contaminated columns
Clean the columns as instructed in the column manual.
•
Plugged Injection valve passages
Rebuild the injection valve (see
4.8
Peak “Ghosting”
Ghosting is the appearance of extraneous peaks in a chromatogram. These may be
late-eluting peaks from a previous injection, or they may result from a
contaminated, malfunctioning, or incorrectly installed injection valve. If these
peaks co-elute with peaks of interest, they will cause nonreproducible peak
heights and/or areas.
•
Insufficient time between sample injections
Wait until the previous sample has been completely eluted before making
another injection.
•
Insufficient flush between samples
Flush the sample loop with at least 10 loop volumes of deionized water or
sample between sample injections.
•
Incorrect or contaminated standards
Remake the standards.
•
Incorrect or contaminated eluent
Remake the eluent (see
•
Malfunctioning injection valve
Contact Technical Support for Dionex products for assistance.
Содержание Dionex Easion
Страница 1: ...Dionex Easion Ion Chromatography System Operator s Manual Document No 155028 Revision 01 August 2020 ...
Страница 34: ...Dionex Easion Operator s Manual 26 Doc 155028 01 08 20 ...
Страница 68: ...Dionex Easion Operator s Manual 60 Doc 155028 01 08 20 ...
Страница 76: ...Dionex Easion Operator s Manual 68 Doc 155028 01 08 20 Figure 5 2 Dionex Easion Flow Schematic ...
Страница 110: ...Dionex Easion Operator s Manual 102 Doc 155028 01 08 20 ...
Страница 118: ...Dionex Easion Operator s Manual 110 Doc 155028 01 08 20 ...