5. Use a 3-mm Allen wrench to
remove the set screw on the
sampling head (see
Figure 5-2. Removing the Set Screw
6. Pull slightly on the sample
out line to gain a small
amount of slack in the line.
Then, slide the tubing out of
the slot in the sampling head
(see
7. Push down on the sampling
head to move the needle into
the carousel a small amount.
Twist the needle to allow the
sampling tip to exit through
the opening in the sensor
block.
Figure 5-3. Sliding the Sample Out Line
Out of the Sampling Head Slot
Sampling
Head
Remove the
set screw
Slide the
line out of
the slot
Sensor Block