Operation
43
Loading the rotor
Maximum loading
If you wish to centrifuge samples that together with
the adapters exceed the maximum permissible load,
you must either reduce the sample volume or calcu-
late the permissible speed n
perm
according to the fol-
lowing formula:
The
Cryofuge
®
5500i GMP
can reach high speeds
implying enormous centrifugal forces. The rotors are
designed in a way having sufficient residual strength
even at the highest permissible speed.
nperm n
max
However, this safety system presupposes that the
maximum permissible load of the rotor is not ex-
ceeded.
Please note the data on the maximum
permissible load and maximum speed
in chapter “Rotors and accessories” on
page 17.
n
perm
= permissible speed
n
max
= maximum speed
maximum permissible load
actual load
Overloading can result in destruc-
tion and severe damage to the cen-
trifuge!