54 | Controller Menu Maps
Thermo Fisher Scientific CryoExtra CE8100 Series
Hot Gas Bypass Sensor Calibration Menus
Bypass probe two point
calibration complete
Place Bypass Probe in ice
water and press ENTER
Wait for Bypass Probe to
stabilize and then press
ENTER
Bypass Temp 0.0
°
C
Calibration Type SINGLE
POINT
Use
to adjust
Press ENTER for next
ENTER
Wait while probe warms to
room temperature
180 secs
Remove Bypass Probe from
LN
2
and press ENTER
Press ENTER for Bypass
Probe Calibration or press
SETUP for next menu
Bypass Calibration
Place Bypass Probe in LN
2
and press ENTER
Calibration aborted
Bypass Calibration
Place Bypass Probe in
LN
2
and press
ENTER
Wait for Bypass Temp to
stabilize and then press
ENTER Bypass Temp -
195.8°C
Bypass probe single point
calibration complete
Wait for Bypass Temp to sta-
bilize and then press ENTER
Bypass Temp-195.8°C
5 Sec
Time out
ENTER (SINGLE)
ESC
ENTER
ENTER
ENTER
180 sec time out
ENTER
ENTER
ENTER
ENTER
ENTER (TWO)