Cryobiological Storage Vessel
2-1
Thermo Scientific
Section 2
General Specifications
* Actual working time may vary due to current atmospheric conditions, container
history, manufacturing tolerances, and any individual patterns of use.
Environmental Conditions
Operating temperature: -29°C to +60°C
10% to 95% relative humidity, non-condensing.
Altitude limit: 2,000 meters.
Storage: -29°C to +60°C; 10% to 95% relative humidity.
Intended for indoor use only.
BioCane 20
BioCane 34
BioCane 47
BioCane 73
Height (Inches)
25.70"
26.60"
26.50"
27.00"
Diameter (Inches)
14.50"
18.20"
20.00"
22.00"
LN2 Capacity
20.50 L
34.80 L
47.40 L
73.0 L
Static Holding Time (Days)
227
193
121
121
Static LN2 Evaporation Rate*
0.1 L/day
0.18 L/day
0.40 L/day
0.60 L/day
Liquid Phase Capacity 1.2 ml
180 vials
720 vials
1512 vials
2016 vials