Installation
Power Wiring
Thermo Scientific
4-3
Figure 3. Power Connections
5. Connect the Ground wire from the main lug on the ground bar at the left of
Terminal Block 1TB to a ground in the facility. Make sure the ground wire
connects to a bare metal surface. Always conform to wiring codes.
6. Install the rear access panel with the screws removed in Step 1.
Wire
1TB
L1
L1
L2
L2
1TB is located include the back of the furnace and on the right
side. (Refer to
Make sure that the furnace is 100% grounded. Do not ground
the furnace to a painted surface. The ground Connection must
be to a clean, bare metal surface to assure proper furnace
operation.